Specifications |
ProLYTE |
| Sample Type |
Serum, Plasma, Whole Blood, Urine |
Sample Size |
100µl Whole Blood, Serum or Plasma
400µl diluted Urine (1:10) |
Analysis Time |
55 sec. (Blood)
90 sec. (Urine) |
| Detection range |
Blood:
Na+: 45-205 mmol/L
K+: 0.5-11 mmol/L
Cl-: 45-205 mmol/L |
Urine:
Na+: 30-1020 mmol/L (undiluted)
K+: 20-505 mmol/L (undiluted)
Cl-: 25-505 mmol/L (undiluted) |
Reproducibility
Typical Within Run (n=20) |
Blood, Serum, Plasma:
Na+: CV ≤ 1% (80-200 mmol/L)
K+: CV ≤ 2% (1-10 mmol/L)
Cl: CV ≤ 2% (80-200 mmol/L) |
Urine:
Na+: CV ≤ 5% (25-1000 mmol/L)
K+: CV ≤ 5% (1-500 mmol/L)
Cl: CV ≤ 5% (25-1000 mmol/L)
|
| Data Management |
125 patient results
20 nromal
QC results
40 abnormal QC results |
| Calibration |
Automatic or On-demand |
| Output |
16 x 2 characters alphanumeric LCD display
40 characters per line printer
RS232 serial port
AutoSampler Port |
| Ambient Conditions |
Room temperature, 15°C~32°C, <80% rel. humidity |
| Power |
100 -115V ~ 50/60Hz 0.8A or
220V 50/60Hz 0.4A
(Factory Set) |
| Size & Weight |
42 cm H x 24 cm W x 20 cm D, 5.8kg |
| Approvals |
CE marked , FDA 510K approved |
| |